Whereas standard ChipShop connectors only accept soft-walled tubing, this set of connectors allows connecting any 1/16" OD rigid tubing to a ChipShop chip.
Suitable for rigid tubing (PTFE, PEEK, etc.) with an OD of 1/16" (1.6 mm). Manufactured in blue TPE, max. pressure of 3.2 bar with a 1/16" OD PEEK tubing.
Plugs to block unused ports are also available in our shop.
In the SensorPlug, the optical carbon dioxide sensor is attached to a Mini Luer plug, for easy integration in microfluidic chips via the corresponding Mini Luer port. This allows for non-invasive measurement of partial pressure of levels CO2 in liquids in millifluidic and microfluidic experiments, without any risk of contamination.
SensorPlugs are delivered beta-irradiated and pre-calibrated for immediate use. The CO2 SensorPlug can be combined with pH and O2 SensorPlugs and measurements can be controlled from one computer via the PreSens Measurement Studio 2 software.
The SensorPlugs can be integrated into fiber sensor integration chips and multi-organ platforms from ChipShop.
The complete setup for PreSens technology includes a meter, a sensor, and relevant accessories including the optic fiber. These components can easily be integrated to any microfluidics and organ-on-chip experiments.

Optical sensors offer several advantages that render them particularly well-suited for controlling small sample volumes:
10x Optical CO2 SensorPlugs
| Measurement range |
1 - 25% CO2 at atmospheric pressure (1013 hPa) 10 - 250 hPa pCO2 8 - 180 mmHg pCO2 |
| Resolution at + 20°C |
± 0.06% CO2 at 2% CO2 ± 0.15% CO2at 6% CO2 ± 0.5 mmHg at 15 mm Hg pCO2 ± 1.2 mmHg at 45 mm Hg pCO2 |
| Accuracy | ± 5% of reading or 0.2% (1.5 mmHg); whichever is higher |
| Measurement temperature range | From + 15 °C to + 45 °C |
| Response time (t90) | < 3 min. for change from 2% to 5% (15 mmHg - 38 mmHg) pCO2 |
| Compatible CO2 meters | CO2-1 SMA with SMA-ST-Adapter |
| Compatibility | Aqueous solutions, pH 4-9 |
| Cross-sensitivity | Optical pCO2 sensors display reduced cross-sensitivity to ionic strength (salinity); acetic acid; SO2; HCl vapors |
| Calibration | Pre-calibrated, re-calibration is possible |
| Storage stability | 12 months provided the sensor material is stored in the dark |
| Plug type | Male Mini Luer fluid connector or customized format on request |

PreSens
Optical pH Sensor - pH SensorPlug (pack of 10)
PreSens
Optical Oxygen Sensor - O2 SensorPlug (pack of 10)
PreSens
Compact Optical CO2 Meter
PreSens
Optical Multichannel Meter - O2/pH/CO2
Bartels Mikrotechnik
CO2 Monitoring Kit
Get the latest updates on new products and upcoming sales
Bartels Mikrotechnik
mp-Lowdriver Pump DriverBartels Mikrotechnik
mp-Lowdriver Pump DriverBartels Mikrotechnik
mp-Lowdriver Pump Driver